논문 기본 정보

자료유형
학술저널
저자정보
유인웅 (서울아산병원 영상의학팀) 정재연 (서울아산병원 영상의학팀) 이관섭 (서울아산병원 영상의학팀)
저널정보
대한디지털의료영상학회 대한디지털의료영상학회논문지 대한디지털의료영상학회논문지 제14권 제2호
발행연도
2012.1
수록면
33 - 38 (6page)

이용수

표지
📌
연구주제
📖
연구배경
🔬
연구방법
🏆
연구결과
AI에게 요청하기
추천
검색

초록· 키워드

오류제보하기
In EVAR procedure using long time C-arm, we studied exposure dose reduction and effeciency through measuring surgical staff's ESD by installing lead curtain operating table next to. The height 3 the spot (50cm, 100cm, 150cm) dose was measured on 2 locations for 600sec in the X-ray radiation considering the surgical staff's movement. To install the curtains, we compared before and after the dose. As a result, it can confirm that dose of the installation former of 50cm height and after is reduced about 75% and 91% in 2 locations. In 100cm height, the reduction of the dose was a bit confirmed. There as to dose value, measured on 150cm height the installation former and after was nearly no change. This research examined the exposure dose about the radiation of the surgical staff during EVAR procedure in which the operation time is the long time. It was implemented in the object that it reduces the radiation exposure. It could confirm the certain effect of the experimental result exposure dose reduction In EVAR procedure using long time C-arm, we studied exposure dose reduction and effeciency through measuring surgical staff's ESD by installing lead curtain operating table next to. The height 3 the spot (50cm, 100cm, 150cm) dose was measured on 2 locations for 600sec in the X-ray radiation considering the surgical staff's movement. To install the curtains, we compared before and after the dose. As a result, it can confirm that dose of the installation former of 50cm height and after is reduced about 75% and 91% in 2 locations. In 100cm height, the reduction of the dose was a bit confirmed. There as to dose value, measured on 150cm height the installation former and after was nearly no change. This research examined the exposure dose about the radiation of the surgical staff during EVAR procedure in which the operation time is the long time. It was implemented in the object that it reduces the radiation exposure. It could confirm the certain effect of the experimental result exposure dose reduction.

목차

등록된 정보가 없습니다.

참고문헌 (0)

참고문헌 신청

이 논문의 저자 정보

최근 본 자료

전체보기

댓글(0)

0