Hans-Ulrich Laasch (Department of Radiology The Christie NHS Foundation Trust Manchester UK) Graham D. Milward (Minnova Medical Foundation C.I.C. Wilmslow UK) Derek W. Edwards (Department of Radiology The Christie NHS Foundation Trust Manchester UK)
저널정보
소화기인터벤션의학회 International Journal of Gastrointestinal Intervention International Journal of Gastrointestinal Intervention 제9권 제3호
발행연도
2020.1
수록면
99 - 105 (7page)

이용수

표지
📌
연구주제
📖
연구배경
🔬
연구방법
🏆
연구결과
AI에게 요청하기
추천
검색

초록· 키워드

오류제보하기
Background: To investigate the expansion force of current colonic stents and to match these to industry standards. Methods: Samples of all colonic stents were requested from manufactures world-wide. Expansion forces were tested with an RX650 compression tool (MSI, Flagstaff, AZ, USA). Measurements were averaged over three cycles of compression and expansion, independently performed at 37°C by specialist engineers of MSI. In parallel, a survey was undertaken on standards, and tests used by manufacturers in their production process. As a labbased study, Institutional Review Board approval was not required. Results: A literature search did not identify any industry standards for testing expansion force or a suggested range for this primary stent function. Median expansion force of all stents was 24.4 N, (35.1 N for braided, 20.7 N for knitted stents) with a vast range from 5.6?130.8 N. Covering braided stents in liquid silicone increased their median force 5.5-fold, separate membranes attached to knitted stents only had a minor effect on expansion force. Five of eight manufacturers replied, describing three different test methods with three different units for expansion force. Conclusion: There are no standards on assessing expansion force, or what the ideal range should be. Consequently, the variation is remarkable, but values are not published, and even if they were, they could not be compared. Consequently, interventionists are unable to discriminate between different stents and to select the most suitable device for their patients, and no recommendation can be made on the ‘best stent’. The industry needs an agreed test standard and an acceptable range of stent forces.

목차

등록된 정보가 없습니다.

참고문헌 (28)

참고문헌 신청

함께 읽어보면 좋을 논문

논문 유사도에 따라 DBpia 가 추천하는 논문입니다. 함께 보면 좋을 연관 논문을 확인해보세요!

이 논문의 저자 정보

최근 본 자료

전체보기

댓글(0)

0