Young-Tak Son (Kyungpook National University) KeunBaDa Son (Kyungpook National University) Kyung-Min Kim (Kyungpook National University) Jae-Mok Lee (Kyungpook National University) Jin-Wook Kim (Kyungpook National University) Wonjae Yu (Kyungpook National University) Kyu-Bok Lee (Kyungpook National University)
저널정보
대한치의학회 Journal of Korean Dental Science Journal of korean dental science Vol.14 No.2
발행연도
2021.12
수록면
69 - 78 (10page)

이용수

표지
📌
연구주제
📖
연구배경
🔬
연구방법
🏆
연구결과
AI에게 요청하기
추천
검색
질문

초록· 키워드

오류제보하기
Purpose: The primary objective of this study was to evaluate the change in the temperature of the adhesive resin in polycrystalline ceramic brackets irradiated using a diode laser at different irradiation energy levels and times.
Materials and Methods: For the measurement of the temperature of the adhesive resin, it was applied at the base of the ceramic bracket, a thermocouple was placed at the center of the base surface, the bracket was placed on prepared resin specimens for light curing, and a laser was irradiated to the center of the bracket slot at 5, 7, and 10 W. For the measurement of the temperatures of the enamel under the bracket and pulp cavity, extracted premolar was fixed to a prepared mold and the ceramic bracket was bonded to the buccal surface of the premolar. The Kruskal–Wallis H test and Friedman test were used for statistical analysis.
Result: At 5 W, the temperature of the adhesive resin did not reach the resin softening temperature of 200°C within 30 seconds. At 7 W, it reached 200°C when the ceramic bracket was irradiated continuously for 28 seconds. At 10 W, it reached 200°C when the ceramic bracket was irradiated continuously for 15 seconds. During laser irradiation, the temperature of the enamel under the bracket increased by over 5°C within 15 seconds.
Conclusion: The use of diode laser irradiation for bracket debonding should be carefully considered because the pulp cavity temperature increases by over 5°C within the irradiation time for resin thermal softening.