Recently, the number of interventional procedures has increased dramatically as an alternative of invasive surgical procedures, and patient radiation exposure is increased greatly. In this study, the patient dose of major interventional procedures nationwide was evaluated and the dose references level guideline for major interventional procedures were suggested and a radiation protection guidelines were created.
The author investigated interventional radiology patient doses in several other countries, assessed accuracy of DAP meters embedded in intervention equipments in domestic country, conducted measurement of patient doses for 13 major interventional procedures with use of Dose Area Product(DAP) meters from 23 hospitals in Korea, and referred to 8,415 cases of domestic data related to interventional procedures by radiation exposure after evaluation the actual effectives of dose reduction variables through phantom test. Finally, dose reference level for major interventional procedures was suggested.
In this study, guidelines for patient doses were 237.7 Gy·cm2 in TACE, 17.3 Gy·cm2 in AVF, 114.1 Gy·cm2 in LE PTA & STENT, 188.5 Gy·cm2 in TFCA, 383.5 Gy·cm2 in Aneurysm Coil, 64.6 Gy·cm2 in PTBD, 64.6 Gy·cm2 in Biliary Stent, 22.4 Gy·cm2 in PCN, 4.3 Gy·cm2 in Hickman, 2.8 Gy·cm2 in Chemo-port, 4.4 Gy·cm2 in Perm-Cather, 17.1 Gy·cm2 in PCD, and 357.9 Gy·cm2 in Vis, EMB.
Dose referenece level acquired in this study is considered to be able to use as minimal guidelines for reducing patient dose in the interventional radiology procedures. For the changes and advances of materials and development of equipments and procedures in the interventional radiology procedures, further studies and monitorings are needed on dose reference level Korean DAP dose conversion factor for the domestic procedures.

목차

서 론 1
연구대상 및 방법 3
1. 연구대상 3
2. 연구방법 4
결 과 15
1. 인터벤션장비 내장형 면적선량계 정확도 평가 15
2. 환자선량평가 18
3. 촬영조건에 따른 팬텀실험 22
고 찰 25
결 론 31
참고문헌 32
ABSTRACT 35