Purpose :
The purpose of this study is evaluating usefulness Bi-presso comparing with conventional sandbag in clinical practice
Material and methods :
The pressure measuring system used in this study was Mark-10"s M5i indicator and MR04 Force sensor (Mark-10, NY, USA). In order to compare the compression force, 0.8 Kg and 1.5 Kg self-made sandbags, Hand pressuring method and Bi-presso (KM, Korea) were measured for compression force. To analyze and compare the measuring data, SPSS Version 21.0 (IBM SPSS Statistics Version 21. Chicago, USA) was used. A questionnaire survey was conducted for 32 patients (15 males and 17 females) applying Bi-presso after Central vein catheterization. In addition, we tried to find out the usefulness of clinical use through questionnaire survey by medical staff.
Results :
■Force compression
The result of the average of compression force 0.8 Kg sand bag: 0.11±0.02 KgF, 1.5 Kg sand bag: 0.038±0.01 KgF, human hand: 0.17±0.02 KgF, in a state of discomfort 0.42±0.05 KgF With Bi-presso applied, it is 0.19±0.02 KgF in a comfortable state and 0.42±0.01 KgF in a state of discomfort. (p <.05)
■Survey of patient group
“Are all 32 patients surveyed with tight arms or discomfort?”, “Are you feeling sore or too much pressure on the questions and pressures in the questionnaire?”, “Do you have any other discomfort when sitting or lying down?”. They said that there was no discomfort in all of the questions.
■Survey of medical staff
Sandbag: 2.36±0.89, Bi-presso: 4.1±0.80. The satisfaction score of Bi-presso was 73.72% higher than that of sandbag. The score on whether the correct position compression is performed is sandbag: 2.4±0.94, Bi-Presso: 3.95±0.88. The score for psychological stability is sandbag: 2.05±0.94, Bi--presso: 4.25±0.85 Posture during hemostasis. The score for changeability is sandbag: 2.15±0.87, Bi-presso: 4.35±0.67 When applied to hypoconscious patients, the score for compression maintenance is sandbag: 1.85±0.93, Bi-presso: 4.1±1.1 infection. The score for risk reduction was significantly higher in the sandbag: 1.75±091, Bi-Presso: 4.55±0.60 in all questionnaire items. (p <.05)
Conclusion :
In conclusion, Bi-presso is considered to be a safe and effective alternative hemostatic device that provides high satisfaction to patients and medical staff because it is possible to press the hemostasis site without risk of infection and bed stability compared to sandbags.